Full STI vs STIPA — When to Use Each Speech Intelligibility Measurement Method

Both Full STI and STIPA are standardized in IEC 60268-16 as valid methods for measuring speech intelligibility in electroacoustic systems. Both are widely used for voice alarm commissioning, PA system testing, and speech intelligibility assessment. Understanding the technical differences and practical implications helps engineers choose the right method for each application — and the right instrument. This guide provides the definitive comparison of Full STI and STIPA from a practical field measurement perspective.

Technical Differences

Full STI (Speech Transmission Index) uses 98 modulation frequencies — 14 per octave band across 7 octave bands (125 Hz to 8 kHz). It measures the modulation transfer function (MTF) of the complete transmission path (electroacoustic system + room acoustics) at each of these 98 frequencies, then applies a weighting matrix per IEC 60268-16 to produce the final STI value. Full STI was traditionally the definitive, most accurate measurement method but required 10–15 minutes per position. The Bedrock Elite i10 completes Full STI in under 60 seconds.

STIPA (STI for Public Address) uses 14 modulation frequencies — 2 per octave band across the same 7 octave bands. It is designed as a faster approximation of Full STI that can be measured in 15–30 seconds. STIPA results are typically very close to Full STI for well-behaved linear systems. The slight accuracy reduction compared to Full STI is generally considered acceptable for routine PA and voice alarm commissioning work.

Accuracy Comparison

When to Use Full STI

When STIPA is Sufficient

Instruments

Full STIPA Guide  ·  Full STI in Under 1 Minute  ·  Request a Quote  ·  Compare Instruments